Digital First Dentistry › Disinfection and decontamination of intraoral scanners

Free verifiable CPD · 1 hour

Disinfection and decontamination of intraoral scanners

About this CPD

Format
Free online self-study, about 1 hour: this page, the required reading below, a five-question check and a short feedback form.
CPD hours
1 hour of verifiable CPD
Aim
To help the dental team decontaminate intraoral scanners safely and consistently between patients.
Objectives
  • Classify the parts of an intraoral scanner as non-critical or semi-critical.
  • Describe why the manufacturer’s instructions for use decide the reprocessing method.
  • List the steps of cleaning and reprocessing a reusable tip.
  • Set up a team routine that makes scanner decontamination reliable.
GDC development outcomes
  • C: Maintenance and development of knowledge and skill within your field of practice.
  • D: Maintenance of skills, behaviours and attitudes which maintain patient confidence in you and the dental professions and put patients’ interests first.
Provider and quality assurance
Provided by Bhavin Associates Ltd (company 07713563). Quality assured by Dr Bhavin Bhatt BDS MFGDP RCS (Eng), GDC 76358, for Bhavin Associates Ltd. See how CPD is quality assured.

Why scanners need their own routine

An intraoral scanner is used in the mouth, often many times a day, and mixes electronic parts that cannot be immersed with tips that touch the mucosa. A clear, written routine protects patients and the equipment.

Classifying the parts

  • Semi-critical: parts that contact mucous membranes, such as the scanner tip or sleeve. These need cleaning followed by sterilisation or high-level disinfection, or are single-use, as the manufacturer specifies.
  • Non-critical: the scanner body, cradle, cables, keyboard and screen. These are cleaned and disinfected with a compatible surface disinfectant, or protected with barriers.

The instructions for use decide

Scanner tips differ. Some are autoclavable for a limited number of cycles, some are designed for high-level disinfection, and some are single-use sleeves. Manufacturers specify the methods, temperatures, times and maximum reuse counts, and some warn against combining methods on the same tip. Using a method that is not in the instructions can damage the optics and may not be effective.

Reprocessing a reusable tip

  1. Remove the tip after each patient, wearing gloves.
  2. Clean it to remove saliva and blood, as the instructions describe. Cleaning comes before disinfection or sterilisation.
  3. Dry, pouch and sterilise, or disinfect, exactly as specified.
  4. Track the number of cycles and replace the tip at its stated limit, or sooner if the window is scratched, fogged or cracked.
  5. Store it so it stays clean until use.

The scanner body and cart

Wipe the body, cradle and touch surfaces with a disinfectant the manufacturer lists as compatible, with the tip removed and any cap fitted. Some disinfectants damage plastics and seals. Use barriers on keyboards and screens that are touched during scanning.

Making it reliable

  • Write a one-page protocol from the instructions for use and keep it near the scanner.
  • Train every person who scans, including nurses who scan as part of the team.
  • Keep enough tips for the day's list so reprocessing is never rushed.
  • Include the scanner in your decontamination audits.

In England, follow HTM 01-05 for decontamination in primary care dental practices alongside the manufacturer's instructions; other countries have equivalent guidance.

Required reading

Read these short pages as part of this CPD:

Doing this as a team? Send it to your nurses and hygienists: WhatsApp · Email

Check your learning

Answer all five questions. You need four correct to continue. You can try again.

1. How is a scanner tip that contacts the mucosa classified?

2. What decides the reprocessing method for a tip?

3. What must happen before disinfection or sterilisation?

4. When should a reusable tip be replaced?

5. Which helps make scanner decontamination reliable?